Maintenance & QA/QC
The Lume requires no reagents or calibration. Biofouling of the optical window is the most common cause of anomalous readings.
🧽 Window Cleaning Procedure
- ▸Power off the sensor before removing the cover
- ▸Remove the sensor from the water
- ▸Twist off the sensor cover counterclockwise
- ▸Wipe the optical window gently with a lint-free cloth
- ▸Do not use abrasive materials — scratching the window degrades performance
- ▸Replace cover and verify seal before redeployment
✅ Field QA/QC Checklist
- ▸Match Lume readings against CBT results — consistent agreement indicates valid data
- ▸Confirm battery is adequately charged before each sampling session
- ▸Leave the sensor submerged for 15–18 minutes to allow readings to stabilise
- ▸Allow at least 15 minutes between different sampling conditions before taking the next reading